The DEPOLOX® Pool Compact Controller is applicable for the measurement of chlorine, pH, ORP and temperature.
Chlorine Calibration
During calibration for free chlorine, a zero point calibration and a measured value calibration (DPD1) must be carried out.
Zero Point Calibration
- Press the symbol with the horizontal lines.
- Press the menu Calibration "Cal".
- Tap the measurement "Chlorine".
- Press the parameter "Zero point".
- Close the ball valve on the sample water inlet.
- Wait until the displayed chlorine value no longer changes.
- Press the "Calibration" button. An input field opens.
- Press "Enter" to save the zero point.
- Open the ball valve on the sample water inlet.
- After zero point calibration, wait at least 2 minutes.
Measuring Value Calibration (DPD)
- Open the sample extraction unit (drain) by approximately 1 turn and extract a specimen of the sample water.
- Determine the content of free chlorine in the sample using a photometer.
- Press the parameter "DPD".
- Press the "Calibration" button. An input field opens.
- Use the input keys to enter the determined value.
- Press "Enter" to save the entry.
This concludes the calibration of free chlorine.
PH Calibration
- Press the symbol with the horizontal lines.
- Press the menu Calibration "Cal".
- Tap the measurement "pH".
- Press the parameter "pH 7".
- Close the sample water inlet and sample water outlet and briefly open the sample extraction unit (drain) to release the pressure. Close the sample extraction unit (drain) again.
- Place the beaker into the bottom clip and fill with buffer solution "pH 7.00" or clamp the bag with buffer solution "pH 7.00" into the bottom clip.
- Unscrew the pH sensor from the cover of the cell body of the flow cell.
- Dip the pH sensor through the top clip at least 2 cm deep into the buffer solution and move slightly until the indicated pH value remains constant.
- Press the "Calibration" button. An input field opens.
- Use the input field to enter the value to be calibrated for the buffer solution.
- Press "Enter" to save the entry.
- Remove the buffer solution "pH 7.00" from the bottom clip.
- Wash the sensor in distilled water to prevent carryover of buffer solution.
- Press the parameter "pH X".
- Place the beaker into the bottom clip and fill it with buffer solution "pH 4.65" or clamp a bag with buffer solution "pH 4.65" into the bottom clip.
- Dip the pH sensor at least 2 cm deep into the buffer solution and move gently until the indicated pH value remains constant.
- Press the "Calibration" button. An input field opens.
- Use the keypad to enter the value to be calibrated.
- Press "Enter" to save the entry.
- Remove the pH sensor from the top clip.
- Screw the pH sensor into the cover of the cell body of the flow cell.
- Open the sample water inlet and outlet again.
This concludes the pH calibration.
ORP Calibration (REDOX)
- Press the symbol with the horizontal lines.
- Press the menu Calibration "Cal".
- Select the "ORP" measurement menu.
- Press the parameter "Cal. value".
- Place the beaker into the bottom clip and fill it with calibration solution "478 mV" or clamp a bag with calibration solution "478 mV" into the bottom clip.
- Close the sample water inlet and sample water outlet and briefly open the sample extraction unit (drain) to release the pressure. Close the sample extraction unit (drain) again.
- Screw the ORP sensor into the cover of the cell body of the flow cell.
- Dip the ORP sensor through the top clip at least 2 cm deep into the calibration solution and move it slightly until the indicated pH value remains constant.
- Press the "Calibration" button. An input field opens.
- Use the keypad to enter the value to be calibrated.
- Press "Enter" to save the entry.
- Remove the ORP sensor from the top clip.
- Screw the ORP sensor into the cover of the cell body of the flow cell.
- Open the sample water inlet and outlet again.
This concludes the ORP calibration.
Temperature Calibration
- Press the symbol with the horizontal lines.
- Press the menu Calibration "Cal".
- Tap the measurement "Temperature".
- Press the parameter "Cal. value".
- Perform comparative temperature measurement.
- Use the keypad to enter the value to be calibrated.
- Press "Enter" to save the entry.
This concludes the temperature calibration.